Antigen Quantification (CDA, PTAU, MBP, SYN)
Upload a fluorescence image or z-stack (Zeiss CZI, TIFF, PNG or JPG), pick the antigen, and quantify. Every layer of a stack is analysed separately.
Goal: Detect and quantify puncta with similar fluorescence characteristics while reducing background noise.
Outline colours. CDA and PTAU: red = counted punctum (the outline follows the punctum's own pixels, its half-maximum footprint, which are the pixels its area and mean intensity are measured over; spots that do not stand out from the tissue around them are not counted); yellow = large bright object (cell-body ring, crescent, blob) that is distinct from the tissue around it - puncta and pieces that are part of it take its colour, it is never counted as a punctum and is measured separately; green = filament (thin elongated fibre), not counted as puncta and measured separately; no outline = background or diffusely brighter tissue. MBP and SYN: red = antigen-positive signal (small spots, whole filaments and, for MBP, broad bright bundles), whose area and intensity are reported; yellow (SYN) = large round granule excluded from the SYN signal; blue = dark regions excluded from the tissue area (unstained cell bodies, holes, empty field). Overlays are 1024 x 1024: open one full screen or download the ZIP to inspect the 1-pixel outlines.
Results per layer
Objects (first 500 shown; type punctum, large_object, filament or positive region)
Train the puncta detector on your gold-standard images: upload images for one puncta antigen (CDA or PTAU), enter the manually verified punctum count for each image (same order, separated by commas) and, for multi-layer files, the layer the counts were made on. The app searches detector settings to match your counts and stores them for that antigen.
Calibration report
Process many images with the current settings for the selected antigen; one result row per layer.
Batch results
Antigen-specific behaviour
- CDA: Identify and quantify puncta with similar fluorescence intensity while minimizing background signal (analysed like PTAU).
- PTAU: Detect and quantify puncta with similar fluorescence characteristics while reducing background noise.
- MBP: Identify MBP fluorescence within cells, including whole filaments, while excluding background signal (area and intensity based).
- SYN: Identify and quantify SYN fluorescence, including small, easily missed puncta, by area and intensity (analysed like MBP).
Layers
Every z-layer of a stack is analysed on its own; layers are never merged, so the spatial organisation along z is kept. Results are reported per layer (with the mean of the layers). A punctum that spans several layers appears in each of them, so layer counts should not be added together.
Noise and texture handling
The shot noise of confocal images grows with intensity, so dim layers and dim regions have less noise than bright ones. The app fits the noise model of each file and transforms the image so the noise is uniform. Detection thresholds are then a fixed number of noise sigmas, where the noise includes the local tissue texture (granular cell bodies get a higher threshold than smooth tissue), measured over the tissue only, so partly empty layers are not flooded with detections.
Puncta antigens (CDA, PTAU)
Denoise, background subtraction that follows broad structures such as cell bodies up to their edges, then Laplacian-of-Gaussian blob detection over a range of punctum sizes. Neighbouring detections are one object unless the brightest path between them dips significantly (a resolved pair of puncta stays two; a rod or a curved rim stays one). Each object is measured over its own pixels: the connected region above half of its peak, so the area and mean intensity are not diluted by background. A punctum must stand out from the tissue around it (1.3 times, 1.7 times for footprints over 40 px). Objects larger or longer than any punctum are large bright objects (yellow) when distinct from the tissue around them (1.5 times; dark cell bodies and empty field are not surroundings), filaments (green) when thin and on a detected line, and otherwise diffusely brighter tissue (no outline). Puncta next to a large object or joined to it by a bright path, and the pieces of a ring around a dark cell body, belong to that object, so one object has one colour; separate dots around a cell body stay puncta. Filaments are found by an oriented line filter that requires a line to be brighter than both of its sides and at least 30 percent brighter than its surroundings; short crosses and stars are rejected.
Area antigens (MBP, SYN)
A tissue mask excludes regions clearly darker than the typical tissue level and empty field (blue). Inside the tissue, positive signal is the union of small spots (the puncta detector's footprints, kept whole, checked like counted puncta), whole filaments from the line filter (MBP: also a long, wide kernel that follows fibres along their length; outlined over the fibre's own pixels at least 15 percent brighter than the tissue around them), and for MBP broad bright regions such as fibre bundles (at least 80 percent above the surrounding tissue baseline) (red). For SYN, large round granules are excluded (yellow). The positive area, its fraction of the tissue, the number of puncta (spots), filament area and length, and intensities are reported per layer. A layer without tissue reports blank fractions.
Calibration (training on your gold standard)
Upload gold-standard images with known manual counts (CDA or PTAU). For multi-layer files enter the layer the counts were made on (one number for all files, or one per file). The app searches detector settings (punctum size and noise-sigma threshold) and keeps the combination that best reproduces your counts. Export the JSON to reuse the calibration.
Input formats
Zeiss CZI (the antigen channel is chosen from the channel metadata, DAPI/Hoechst channels are skipped), 8/16-bit grayscale or RGB TIFF (single images, z-stacks, ImageJ hyperstacks; multichannel TIFFs use channel names when present, otherwise channel 1), PNG, JPG. In merged RGB exports the blue (nuclear counterstain) channel is ignored; RGB exports are display-scaled, so their intensities are not comparable with raw CZI values. Nearly black outermost scan lines are repaired. All measurements are in pixels.